Numerical Study of Coupling between Coagulators and Electrodes of Cardiac Pacemakers under Consideration of the Human Body
Keywords:
Numerical Study of Coupling between Coagulators and Electrodes of Cardiac Pacemakers under Consideration of the Human BodyAbstract
The paper presents a numerical, worst-case study of the coupling between coagulators and cardiac pacemakers under consideration of the human body. For this purpose two special body models have been implemented. With these models whose electrical properties correspond to a weighted average of those of different tissues of the human body the influence of the latter can be taken into account in the computations. Different parameters such as position of the pacemaker and its electrode, coagulation frequency, and coagulation point on the surface of the human body are considered. Based on results of previous investigations simplified approaches to account for the dielectric coating of the coagulator electrode and housing of the pacemaker can be used. The investigation shows, that for the scenario described above, strong peaks occur in the resulting graphs arising form resonance effects on the coagulator cable
Downloads
References
Deutsche Industrie Norm, Körpermaße
des Menschen; Werte, DIN 33 402 Teil 2,
S. 1–33, Okt. 1986.
Deutsche Industrie Norm, Körpermaße
des Menschen; Werte; Anwendung von
Körpermaßen in der Praxis, Beiblatt zu
DIN 33 402 Teil 2, S. 1–5, Okt. 1984.
EM Software & Systems, FEKO User’s
Manual, Suite 3.2.1, March 2002,
www.feko.co.za.
C. Gabriel, “Compilation of the Dielectric
Properties of Body Tissues at RF and Mi-
crowave Frequencies,” Brooks Air Force
Technical Report AL/OE-TR-1996-0037,
Jun. 1996.
Italian National Research Council, Di-
electric Properties of Body Tissues in the
frequency range 10 Hz – 100 GHz, Elec-
tromagnetic Wave Research Institute,
Florence Italy, Available at
http://safeemf.iroe.fi.cnr.it/tissprop/.
U. Jakobus, “Erweiterte Momenten-
methode zur Behandlung kompliziert
aufgebauter und elektrische großer elekt-
romagnetischer Streuprobleme,“ Disserta-
tion, Nr. 171 in Fortschrittsberichte, Rei-
he 21, Düsseldorf: VDI Verlag, 1995.
J. G. Koritke and H. Sick, Atlas of Sec-
tional Human Anatomy, Baltimore: Urban
& Schwarzenberg, 1988.
ACES JOURNAL, VOL. 19, NO. 1b, MARCH 2004
F. Landstorfer et al., Development of a
model describing the coupling between
electrodes of cardiac pacemakers and
transmitting antennas in their close vicin-
ity in the frequency range from 50 MHz to
MHz“, Study for the Forschungsge-
meinschaft Funk e. V., Institut für
Hochfrequenztechnik, Universität Stutt-
gart 1999.
M. Schick, ”Numerical Study of Cou-
pling between Coagulators and Electrodes
of Cardiac Pacemakers,“ Proceedings of
the 31st European Microwave Conference
EuMC 2001, vol. 3, pp. 231–234, Lon-
don, Sep. 2001