Numerical Study of Coupling between Coagulators and Electrodes of Cardiac Pacemakers under Consideration of the Human Body

作者

  • Markus Schick Institut fuer Hochfrequenztechnik, Universitaet Stuttgart Pfaffenwaldring 47, 70550 Stuttgart, Germany
  • Friedrich M. Landstorfer Institut fuer Hochfrequenztechnik, Universitaet Stuttgart Pfaffenwaldring 47, 70550 Stuttgart, Germany

关键词:

Numerical Study of Coupling between Coagulators and Electrodes of Cardiac Pacemakers under Consideration of the Human Body

摘要

The paper presents a numerical, worst-case study of the coupling between coagulators and cardiac pacemakers under consideration of the human body. For this purpose two special body models have been implemented. With these models whose electrical properties correspond to a weighted average of those of different tissues of the human body the influence of the latter can be taken into account in the computations. Different parameters such as position of the pacemaker and its electrode, coagulation frequency, and coagulation point on the surface of the human body are considered. Based on results of previous investigations simplified approaches to account for the dielectric coating of the coagulator electrode and housing of the pacemaker can be used. The investigation shows, that for the scenario described above, strong peaks occur in the resulting graphs arising form resonance effects on the coagulator cable

##plugins.generic.usageStats.downloads##

##plugins.generic.usageStats.noStats##

参考

Deutsche Industrie Norm, Körpermaße

des Menschen; Werte, DIN 33 402 Teil 2,

S. 1–33, Okt. 1986.

Deutsche Industrie Norm, Körpermaße

des Menschen; Werte; Anwendung von

Körpermaßen in der Praxis, Beiblatt zu

DIN 33 402 Teil 2, S. 1–5, Okt. 1984.

EM Software & Systems, FEKO User’s

Manual, Suite 3.2.1, March 2002,

www.feko.co.za.

C. Gabriel, “Compilation of the Dielectric

Properties of Body Tissues at RF and Mi-

crowave Frequencies,” Brooks Air Force

Technical Report AL/OE-TR-1996-0037,

Jun. 1996.

Italian National Research Council, Di-

electric Properties of Body Tissues in the

frequency range 10 Hz – 100 GHz, Elec-

tromagnetic Wave Research Institute,

Florence Italy, Available at

http://safeemf.iroe.fi.cnr.it/tissprop/.

U. Jakobus, “Erweiterte Momenten-

methode zur Behandlung kompliziert

aufgebauter und elektrische großer elekt-

romagnetischer Streuprobleme,“ Disserta-

tion, Nr. 171 in Fortschrittsberichte, Rei-

he 21, Düsseldorf: VDI Verlag, 1995.

J. G. Koritke and H. Sick, Atlas of Sec-

tional Human Anatomy, Baltimore: Urban

& Schwarzenberg, 1988.

ACES JOURNAL, VOL. 19, NO. 1b, MARCH 2004

F. Landstorfer et al., Development of a

model describing the coupling between

electrodes of cardiac pacemakers and

transmitting antennas in their close vicin-

ity in the frequency range from 50 MHz to

MHz“, Study for the Forschungsge-

meinschaft Funk e. V., Institut für

Hochfrequenztechnik, Universität Stutt-

gart 1999.

M. Schick, ”Numerical Study of Cou-

pling between Coagulators and Electrodes

of Cardiac Pacemakers,“ Proceedings of

the 31st European Microwave Conference

EuMC 2001, vol. 3, pp. 231–234, Lon-

don, Sep. 2001

##submission.downloads##

已出版

2022-06-18

栏目

General Submission